TECHNICAL DETAILS

General technical data

Dimensions (W/H/D)
mm
760 x 1230 x 730
Weight
kg
230

Electrical supply data

Nominal voltage (AC)
VAC
230 ± 10
Supply frequency
Hz
50 - 60
Maximum intake current
A
12
Fuse protection provided by customer
A
Connect to a separate power circuit with residual circuit breaker B16A
Protection rating
-
IP 20 (Device is protected against access of objects, but not protected against access of water)

Environmental conditions for storage and operation

Temperature range
°C
5 - 40
Relative humidity, maximum
% rH
80
Device positioning
-
Dry interiors
Operating altitude
m AMSL (max.)
2.000
Pollution degree
m AMSL (max.)
2

Incubator

Internal volume
L
240
Setting temperature range
°C
+18 to +50
Setting temperature step
°C
0.1
Setting relative humidity range
% rH
30 - 90
Setting O2 range
%
1 - 21

Robot

Dimensions (W/H/D)
mm
420 x 360 x 360
Weight
kg
8.5
Pipetting volume
µl
5 - 200
Radial operation range
mm
180
Horizontal positioning accuracy
µm
± 35
Vertical positioning accuracy
µm
± 5
Maximum movement speed
mm/s
470
Positioning time
s
< 1.8
Maximum amount of pipette tips
24
Temperature range
°C
5 - 40
Relative humidity
% rH
10 – 80 (non condensing)
Nominal voltage (DC)
VDC
24

Microscope imaging system

Dimensions of body (W/H/D)
mm
100 x 145 x 156.5
Objectives
-
1
Objective magnification
X
20
Z – drive type
-
VoiceCoil
Focus range / resolution
µm
6800 / 0.002
Nominal voltage (DC)
VDC
+-15
Maximum current consumption
A
1
Camera resolution
MPx
12.3
Pixel size
µm
3.45

Cellular impedance

Throughput (wells per second)
1 / s
14
Measuring frequency
kHz
10
Measuring range impedance Z
10 - 5000
Measuring range resistance R
10 - 5000
Measuring range capacity C
nF
0.3 - 3000
Nominal voltage
mV
20
Current density
mA / cm²
< 2
Measuring error
%
< 2
Resolution impedance magnitude
0.01
Resolution impedance phase
°
0.01
Frequency accuracy
%
< 0.01
Current range
mA
0.01 - 10
Reproducibility (well to well)
%
< 2
Sensitivity
%
0.01
Response time
s
0.01
Signal drift
% / d
0.1
Accurateness
%
< 5
Reproducibility (well to well)
%
< 2

pH and pO2 single channel measuring device

Measuring range
pO2
pH
Measuring range
0 - 50 %
6.0 - 8.5 pH
Resolution
± 0.4 % pO2
± 0.05 pH
Response time
s
< 30
Signal drift
s
0.2 % / week
0.1 pH / week
Accurateness
± 1.0 % at 20.9 % pO2
± 0.2 % at pH7
Reproducibility (well to well)
%
< 5

CYRIS® FLOX technical data sheet